Fiveable

🥼Organic Chemistry Unit 12 Review

QR code for Organic Chemistry practice questions

12.5 Spectroscopy and the Electromagnetic Spectrum

12.5 Spectroscopy and the Electromagnetic Spectrum

Written by the Fiveable Content Team • Last updated August 2025
Written by the Fiveable Content Team • Last updated August 2025
🥼Organic Chemistry
Unit & Topic Study Guides

Electromagnetic radiation interacts with molecules in specific, predictable ways. Light can cause bonds to vibrate, molecules to rotate, or electrons to jump between energy levels. These interactions are the foundation of spectroscopy, which you'll use throughout organic chemistry to figure out what molecules look like.

Different types of radiation excite different molecular motions. Infrared light causes bond vibrations, while UV and visible light trigger electronic transitions. By studying which wavelengths a molecule absorbs, you can work backward to deduce its structure and functional groups.

Electromagnetic Radiation and Spectroscopy

Electromagnetic radiation and organic molecules

Electromagnetic radiation consists of oscillating electric and magnetic fields traveling through space as waves. Three properties characterize these waves:

  • Wavelength (λ\lambda): the distance between consecutive wave peaks
  • Frequency (ν\nu): the number of wave cycles per second (measured in Hz)
  • Energy (EE): directly proportional to frequency, inversely proportional to wavelength

A molecule absorbs a photon when the photon's energy exactly matches the energy gap between two of the molecule's energy levels. This causes the molecule to jump from a lower energy state to a higher one.

Different molecular motions require different amounts of energy:

  • Bond vibrations (stretching and bending) absorb infrared (IR) radiation
  • Electronic transitions (electrons moving between orbitals) absorb ultraviolet (UV) and visible light, which is higher in energy than IR

Wavelengths that aren't absorbed simply pass through the sample. That's why some materials appear transparent to certain types of radiation.

Electromagnetic radiation and organic molecules, 6.1 Electromagnetic Energy – Chemistry

Energy levels and transitions

Molecules have discrete (quantized) energy levels, meaning they can only exist in specific energy states:

  • Ground state: the lowest energy configuration
  • Excited states: higher energy configurations the molecule can be promoted to

Two key processes connect these levels:

  • Absorption: the molecule takes in a photon and moves to a higher energy level
  • Emission: the molecule releases a photon and drops back to a lower energy level

Electronic transitions involve redistributing electrons among molecular orbitals. These transitions require significantly more energy than vibrational transitions, which is why electronic spectroscopy uses UV/visible light while vibrational spectroscopy uses lower-energy IR light.

Electromagnetic radiation and organic molecules, Infrared spectroscopy - Wikipedia

Photon energy calculations

The Planck equation relates a photon's energy to its frequency and wavelength:

E=hν=hcλE = h\nu = \frac{hc}{\lambda}

  • EE = energy of the photon (joules)
  • hh = Planck's constant (6.626×10346.626 \times 10^{-34} J·s)
  • ν\nu = frequency (Hz)
  • cc = speed of light (2.998×1082.998 \times 10^{8} m/s)
  • λ\lambda = wavelength (meters)

The key relationship to internalize: frequency and wavelength are inversely proportional. Higher frequency means shorter wavelength and higher energy. Lower frequency means longer wavelength and lower energy.

The electromagnetic spectrum, ordered from lowest to highest energy:

  1. Radio waves — longest wavelengths, lowest energies
  2. Microwaves — used in ovens; cause molecular rotations
  3. Infrared (IR) — causes bond vibrations; the basis of IR spectroscopy
  4. Visible light — the narrow range your eyes detect
  5. Ultraviolet (UV) — causes electronic transitions; used in UV-Vis spectroscopy
  6. X-rays — high enough energy to interact with inner-shell electrons
  7. Gamma rays — shortest wavelengths, highest energies; from nuclear processes

For organic chemistry, you'll mostly work in the IR and UV-Vis regions.

Interpretation of infrared spectra

IR spectroscopy measures how much infrared radiation a sample absorbs. The result is plotted as percent transmittance (%T) on the y-axis versus wavenumber (ν~\tilde{\nu}) on the x-axis.

Wavenumber is the reciprocal of wavelength, expressed in cm1\text{cm}^{-1}. Higher wavenumber means higher energy. This is the standard x-axis unit in IR spectroscopy, and the axis runs from high wavenumber on the left to low wavenumber on the right.

Reading the spectrum takes some practice. Absorption peaks point downward (toward lower %T), because less light is transmitted at wavelengths where the molecule absorbs. A dip in the spectrum means the molecule is absorbing at that wavenumber.

Different functional groups absorb at characteristic wavenumber ranges. The most important ones to memorize:

  • O–H and N–H stretches: 3200–3600 cm1\text{cm}^{-1} (broad for O–H in alcohols and carboxylic acids; medium for N–H in amines)
  • C–H stretches: 2800–3300 cm1\text{cm}^{-1} (nearly every organic molecule shows these)
  • C≡C and C≡N stretches: ~2100–2300 cm1\text{cm}^{-1} (alkynes and nitriles)
  • C=O stretches: 1650–1800 cm1\text{cm}^{-1} (strong, sharp peaks for ketones, aldehydes, esters, carboxylic acids)
  • C=C stretches: 1600–1680 cm1\text{cm}^{-1} (alkenes; often weaker than C=O)

When analyzing an IR spectrum, you're looking for two things: which peaks are present and which are absent. The presence of a broad absorption around 3300 cm1\text{cm}^{-1} suggests an O–H or N–H group. The absence of a strong peak near 1700 cm1\text{cm}^{-1} tells you there's likely no carbonyl group.

You can also compare an unknown compound's spectrum against reference spectra. The region below ~1500 cm1\text{cm}^{-1} is called the fingerprint region because its complex pattern of peaks is unique to each molecule, much like an actual fingerprint.

The Beer-Lambert law (A=εbcA = \varepsilon bc) relates absorbance to the concentration of a sample (cc), the path length of the sample cell (bb), and the molar absorptivity (ε\varepsilon). While you'll encounter this more in UV-Vis spectroscopy, it's the principle that allows spectroscopy to be used for quantitative analysis, not just identification.