Sound level meters are crucial tools in architectural acoustics, measuring and quantifying sound pressure levels. They come in different types, including Class 1 and Class 2 meters, as well as integrating and non-integrating models, each designed for specific accuracy requirements and measurement capabilities.
These devices consist of key components like microphones, preamplifiers, frequency weighting networks, and RMS detectors. Understanding their proper use, calibration, and maintenance is essential for obtaining accurate and reliable measurements in various architectural acoustic applications.
Types of sound level meters
Sound level meters are essential tools for measuring and quantifying sound pressure levels in architectural acoustics
Different types of sound level meters are designed to meet specific accuracy requirements and measurement capabilities
Class 1 vs class 2 meters
Top images from around the web for Class 1 vs class 2 meters
Acoustic quantities, part 1: What are decibels? - Erlend M. Viggen View original
Is this image relevant?
1 of 3
Class 1 sound level meters offer the highest accuracy and are used for precise measurements in critical applications (laboratory testing, legal metrology)
Conform to strict tolerance limits for , linearity, and environmental influences
Typically more expensive than Class 2 meters
Class 2 sound level meters provide a balance between accuracy and cost-effectiveness
Suitable for general-purpose acoustic measurements and noise assessments
Meet slightly more relaxed tolerance limits compared to Class 1 meters
Integrating vs non-integrating meters
Integrating sound level meters can calculate and display time-averaged sound levels (Leq, Lden)
Continuously measure and integrate sound energy over a specified time period
Useful for assessing long-term noise exposure and compliance with noise regulations
Non-integrating meters, also known as conventional sound level meters, provide instantaneous sound pressure level readings
Display the current sound level at any given moment
Suitable for spot-checking noise levels and identifying peak sound events
Components of sound level meters
Sound level meters consist of several key components that work together to accurately measure and process sound pressure levels
Microphone and preamplifier
The microphone converts sound pressure variations into electrical signals
Condenser microphones are commonly used due to their stability and wide frequency response
The microphone's sensitivity and frequency response influence the meter's overall accuracy
The preamplifier amplifies the weak electrical signal from the microphone
Provides impedance matching and signal conditioning
Minimizes noise interference and ensures a clean signal for further processing
Frequency weighting networks
Frequency weighting networks filter the sound signal to mimic human hearing sensitivity at different frequencies
(dBA) is the most common, emphasizing frequencies around 1-6 kHz
C-weighting (dBC) provides a flatter response, suitable for low-frequency noise and peak measurements
Z-weighting (dBZ) applies no frequency weighting, representing the unfiltered sound pressure level
The selected frequency weighting affects the displayed sound level values and interpretation of results
RMS detector and time averaging
The RMS (Root Mean Square) detector calculates the effective sound pressure level from the frequency-weighted signal
Provides a measure of the average sound energy over time
Smooths out fluctuations and gives a stable reading
Time averaging constants determine the meter's response to changes in sound level
Fast (125 ms), Slow (1 s), and Impulse (35 ms) time constants are available
The choice of time constant depends on the nature of the sound being measured (steady-state, fluctuating, or impulsive)
Display and data logging
The display shows the measured sound level in decibels (dB) and other relevant information (frequency weighting, time averaging, overload indicators)
Digital displays offer high resolution and clarity
Analog displays with needles provide a visual representation of sound level variations
Data logging capabilities allow for the storage and retrieval of measurement data
Built-in memory or external storage devices (SD cards) can be used
Logged data can be transferred to a computer for further analysis and reporting
Calibration of sound level meters
Regular calibration of sound level meters is crucial to ensure accurate and reliable measurements
Acoustic calibrator usage
An acoustic calibrator generates a known sound pressure level at a specific frequency (typically 1 kHz at 94 dB or 114 dB)
The calibrator is fitted over the microphone, creating a sealed coupling
The meter's reading is adjusted to match the calibrator's output level
Calibration checks should be performed before and after each measurement session
Field calibration procedures
involves checking and adjusting the meter's sensitivity in the actual measurement environment
Accounts for the influence of temperature, humidity, and atmospheric pressure on the microphone's response
Ensures the meter is accurately measuring sound levels under the prevailing conditions
Follow the manufacturer's recommended procedure for field calibration
Calibration documentation and traceability
Calibration certificates and records should be maintained for each sound level meter
Document the calibration date, reference standards used, and any adjustments made
Ensure traceability to national or international standards (NIST, ISO)
Regular factory calibration by an accredited laboratory is recommended (typically annually)
Verifies the meter's compliance with specifications and performance standards
Provides independent validation of the meter's accuracy and reliability
Measurement settings and parameters
Proper selection of measurement settings and parameters is essential for obtaining accurate and meaningful results
Frequency weighting (A, C, Z)
Choose the appropriate frequency weighting based on the type of noise being measured and the relevant standards or regulations
A-weighting (dBA) is commonly used for environmental noise assessments and hearing protection purposes
C-weighting (dBC) is used for low-frequency noise, sound insulation testing, and peak level measurements
Z-weighting (dBZ) is used when a flat, unweighted response is required
Ensure the meter is set to the correct frequency weighting before starting measurements
Time weighting (Fast, Slow, Impulse)
Select the appropriate time weighting constant based on the characteristics of the sound being measured
Fast (125 ms) is suitable for most general-purpose measurements and fluctuating noise
Slow (1 s) is used for more stable, average sound levels and assessing human response to noise
Impulse (35 ms) is used for measuring short-duration, impulsive sounds (gunshots, impacts)
The time weighting affects the meter's ability to capture and display rapid changes in sound level
Measurement range and overload
Set the measurement range to accommodate the expected sound levels without overloading the meter
Overload occurs when the sound level exceeds the meter's maximum measurable limit
An overload indicator alerts the user to adjust the range or use a microphone with a higher
Use a suitable measurement range to optimize the meter's resolution and avoid missing important data
Peak vs RMS measurements
Peak measurements capture the instantaneous maximum sound pressure level
Used for assessing short-duration, impulsive sounds and determining the crest factor
Peak levels are typically higher than RMS levels and do not represent the average sound energy
RMS measurements provide the effective sound pressure level over time
Used for most general-purpose acoustic measurements and noise assessments
RMS levels are more representative of the perceived loudness and potential hearing damage risk
Measurement procedures and techniques
Following standardized measurement procedures and techniques ensures consistent and reliable results
Microphone placement and orientation
Position the microphone at the appropriate distance and orientation relative to the sound source
For free-field measurements, the microphone should be pointed directly at the source
For diffuse-field measurements, the microphone should be oriented at a 70-80° angle to minimize reflections
Consider the microphone's height, proximity to reflective surfaces, and potential interference from the operator's body
Background noise considerations
Assess the background noise level before conducting measurements
Background noise can mask the sound of interest and affect the accuracy of results
Use a windscreen to reduce wind noise and protect the microphone from dust and moisture
If possible, measure the sound source with and without the presence of background noise for comparison
Measurement duration and sampling
Determine the appropriate measurement duration based on the type of sound and the purpose of the assessment
Steady-state sounds require shorter measurement times (e.g., 30 seconds) to obtain representative levels
Fluctuating or intermittent sounds may require longer durations (e.g., 15 minutes) to capture variations over time
Use an adequate sampling rate to ensure accurate representation of the sound signal
The sampling rate should be at least twice the highest frequency of interest (Nyquist criterion)
Higher sampling rates provide better time resolution but generate larger data files
Spatial averaging and sound field characterization
Conduct measurements at multiple positions to account for spatial variations in the sound field
Use a grid or pattern to systematically cover the area of interest
Measure at ear height for assessments related to human exposure and perception
Characterize the sound field as free-field, diffuse-field, or near-field based on the measurement results
Free-field conditions exist when the sound propagates without reflections (outdoor measurements)
Diffuse-field conditions occur in reverberant spaces with sound coming from all directions (indoor measurements)
Near-field measurements are taken close to the sound source, where the sound pressure level varies significantly with distance
Analysis and interpretation of results
Proper analysis and interpretation of sound level measurement results are crucial for drawing meaningful conclusions and making informed decisions
Equivalent continuous sound level (Leq)
Leq represents the average sound level over a specified time period, expressed in decibels (dB)
Calculated by integrating the sound energy over time and dividing by the measurement duration
Provides a single-number representation of the time-varying sound level
Leq is commonly used for assessing long-term noise exposure, community noise, and compliance with noise regulations
Statistical noise levels (L10, L50, L90)
Statistical noise levels describe the percentage of time that a certain sound level is exceeded
L10 represents the sound level exceeded for 10% of the measurement time, indicating the higher end of the noise range
L50 represents the median sound level, exceeded for 50% of the time
L90 represents the background noise level, exceeded for 90% of the time
Statistical levels provide insight into the variability and character of the noise environment
Noise dose and exposure calculations
Noise dose is a measure of the total sound energy a person is exposed to over a specific period (usually 8 hours)
Calculated based on the average sound level and the exposure duration
Expressed as a percentage of the maximum allowable daily dose (100% dose equals the permissible exposure limit)
Noise exposure calculations help assess the potential risk of hearing damage and guide the selection of appropriate hearing protection measures
Comparison to noise criteria and regulations
Compare the measured sound levels to relevant noise criteria, standards, and regulations
Noise criteria (NC) curves define acceptable noise levels for different building types and room functions
Standards (ANSI, ISO) provide guidelines for measuring and evaluating noise in specific environments
Regulations (OSHA, EPA) set limits on noise exposure to protect public health and worker safety
Identify areas of non-compliance and develop strategies for noise control and mitigation
Maintenance and care of sound level meters
Proper maintenance and care of sound level meters are essential for ensuring their long-term performance and reliability
Handling and storage guidelines
Handle the sound level meter with care to avoid physical damage
Use a protective case or pouch for transportation and storage
Avoid exposing the meter to extreme temperatures, humidity, or dust
Store the meter in a clean, dry environment when not in use
Keep the meter away from direct sunlight and heat sources
Use desiccants to control moisture in the storage case
Battery management and replacement
Regularly check the battery level and replace batteries as needed
Low battery voltage can affect the meter's accuracy and cause erratic readings
Use high-quality, fresh batteries of the type recommended by the manufacturer
Remove batteries from the meter if it will not be used for an extended period to prevent leakage and corrosion
Cleaning and protection of microphone
Keep the microphone clean and free from debris, dust, and moisture
Use a soft brush or compressed air to gently remove particles from the microphone surface
Avoid touching the microphone diaphragm with fingers or sharp objects
Use a windscreen or foam cover to protect the microphone during outdoor measurements
Windscreens reduce wind noise and prevent dust and moisture from entering the microphone
Replace worn or damaged windscreens to maintain optimal performance
Regular functionality checks and testing
Perform regular functionality checks to ensure the meter is operating correctly
Verify the meter's response to a known sound source (e.g., acoustic calibrator)
Check the battery voltage, display, and data logging functions
Schedule periodic testing and recalibration by a qualified service center
Identify and correct any deviations from the manufacturer's specifications
Ensure the meter continues to meet the required accuracy and performance standards
Applications of sound level meters in architectural acoustics
Sound level meters are versatile tools with numerous applications in the field of architectural acoustics
Room acoustics measurements
Assess the acoustic properties of rooms and spaces using sound level meters
Measure reverberation time (RT) by recording the sound decay after a broadband noise source is turned off
Evaluate speech intelligibility by measuring the speech transmission index (STI) or clarity index (C50)
Identify room modes, flutter echoes, and other acoustic defects through detailed measurements
Use the results to optimize room acoustics through the selection of appropriate materials, surfaces, and geometries
Noise control and mitigation
Investigate and quantify noise sources in buildings using sound level meters
Identify the dominant noise contributors (HVAC systems, plumbing, exterior sources)
Measure noise levels in different areas and compare them to acceptable criteria
Develop noise control strategies based on the measurement results
Implement sound absorption, sound barriers, vibration isolation, or active noise control
Verify the effectiveness of noise mitigation measures through post-treatment measurements
Building code compliance testing
Use sound level meters to demonstrate compliance with building codes and acoustic performance standards
Measure airborne and impact sound insulation between rooms or units (STC, IIC ratings)
Assess the background noise levels in different room types and compare them to the specified criteria
Verify the sound power levels of building equipment and appliances
Provide measurement reports and documentation to certify compliance and obtain occupancy permits
Environmental noise assessment
Evaluate the impact of environmental noise on buildings and their occupants using sound level meters
Measure outdoor noise levels from traffic, aircraft, construction, or industrial sources
Assess the noise exposure at different facades and elevations of the building
Predict indoor noise levels based on the measured outdoor levels and the building envelope performance
Recommend appropriate façade designs, window treatments, or site planning strategies to mitigate environmental noise
Key Terms to Review (18)
A-weighting: A-weighting is a frequency weighting system that adjusts the relative loudness of different frequencies to reflect human hearing sensitivity. It is commonly used in sound measurement to ensure that noise levels are reported in a way that corresponds to how we perceive sound, making it relevant for evaluating various acoustic environments and noise impacts.
ANSI Standards: ANSI standards are guidelines and specifications developed by the American National Standards Institute to ensure quality, safety, and efficiency in various industries. These standards play a critical role in establishing uniformity in measurements, materials, and procedures, which are essential for effective communication and performance in fields like architectural acoustics.
Decibel (dB): A decibel (dB) is a logarithmic unit used to measure sound level, indicating the intensity of sound relative to a reference level. It provides a way to express the ratio of two values, such as power or intensity, in a compact form, making it easier to compare sound levels. The decibel scale is particularly useful because human perception of sound is logarithmic, meaning we perceive changes in sound intensity differently at different levels.
Dynamic Range: Dynamic range refers to the difference between the smallest and largest possible values of a signal, particularly in terms of sound intensity. This concept is crucial in various fields as it determines how much variation in loudness can be captured or reproduced without distortion. Understanding dynamic range is vital for accurately measuring sound levels, analyzing frequency content, and processing audio signals, ensuring fidelity in acoustic environments.
Environmental noise monitoring: Environmental noise monitoring refers to the systematic measurement and assessment of noise levels in a specific environment to evaluate its impact on the community and surroundings. This process involves using specialized equipment, like sound level meters, to collect data on noise pollution over time, helping to inform regulations, urban planning, and public health initiatives.
EPA Noise Guidelines: EPA Noise Guidelines refer to the standards and recommendations set forth by the Environmental Protection Agency (EPA) to manage and mitigate noise pollution. These guidelines provide a framework for assessing noise levels, determining acceptable exposure limits, and identifying strategies for reducing noise in various environments. They play a crucial role in protecting public health and welfare, especially in urban settings where noise can significantly impact quality of life.
Equivalent Continuous Sound Level: Equivalent continuous sound level, often abbreviated as Leq, is a measure of the continuous sound level that, over a specified period, has the same energy as the fluctuating sound levels during that time. This concept is crucial for understanding how varying sound levels can be averaged into a single value that represents overall exposure to sound, making it especially relevant in evaluating noise environments using sound level meters. The Leq provides a useful metric for assessing the impact of noise in various settings by incorporating both the intensity and duration of sound events.
Field Calibration: Field calibration is the process of adjusting and verifying the accuracy of sound level meters in real-world environments to ensure they provide precise measurements. This process often involves comparing the readings from the sound level meter with a known reference or standard and making necessary adjustments to align them. Field calibration is crucial for maintaining the integrity of acoustic measurements in various settings, such as construction sites, concert venues, and noise monitoring programs.
Frequency Response: Frequency response refers to the measure of an audio system's output spectrum in response to an input signal across a range of frequencies. It reflects how different frequencies are amplified or attenuated by a system, impacting sound clarity and quality in various acoustic environments.
Humidity influence: Humidity influence refers to the effect that moisture content in the air has on sound propagation and acoustical measurements. As humidity levels change, they can alter sound speed, absorption, and overall sound quality, which is crucial when using sound level meters for accurate readings in various environments.
IEC 61672: IEC 61672 is an international standard that specifies the performance requirements for sound level meters, including their design, construction, and measurement capabilities. This standard is crucial as it ensures consistency and accuracy in sound level measurements across different devices, which is essential for assessing environmental noise and ensuring compliance with regulations.
Integrating sound level meter: An integrating sound level meter is a specialized device used to measure sound levels over a specified time period, providing an average sound level as well as peak levels. This type of meter is essential for assessing noise exposure in various environments, allowing users to understand both continuous and fluctuating sounds more effectively. It typically incorporates features such as frequency weighting and time averaging to accurately reflect the impact of noise on human hearing.
Laboratory calibration: Laboratory calibration is the process of adjusting and verifying the accuracy of measuring instruments, such as sound level meters, by comparing their output against a known standard in a controlled environment. This ensures that the instruments provide reliable and precise measurements, which are critical for various applications, including sound measurement in architectural acoustics. Proper calibration helps to maintain the integrity of data collected and ensures compliance with industry standards.
Local noise ordinances: Local noise ordinances are regulations established by municipalities to control and limit excessive noise levels in specific areas to protect public health, safety, and general welfare. These rules are often based on specific time periods and types of noise, and they play a crucial role in setting acceptable sound levels, which relate directly to noise criteria and ratings, as well as the measurement of sound using sound level meters.
Noise dosimeter: A noise dosimeter is a specialized device used to measure an individual's exposure to sound levels over time, providing a comprehensive assessment of noise exposure in various environments. It records sound levels in decibels (dB) and integrates this data to calculate the total noise exposure, which is particularly important in occupational settings to ensure compliance with safety regulations. This tool helps in understanding potential hearing risks and implementing appropriate hearing conservation measures.
Occupational Health: Occupational health refers to the branch of public health that focuses on the physical and mental well-being of workers in their workplace environments. It involves the identification, evaluation, and control of hazards that could lead to injuries or illnesses due to workplace exposure. This field aims to promote safe work practices and ensure a healthy work environment, particularly in relation to exposure to noise, chemicals, and other occupational risks.
Peak Sound Level: Peak sound level is the maximum instantaneous sound pressure level measured in decibels (dB) during a specific period. This measurement captures the loudest point of a sound wave, providing essential information about transient sounds and their potential impact on hearing and environmental noise. Peak sound levels are critical for assessing loudness in various contexts, including construction sites, concerts, and industrial environments.
Temperature correction: Temperature correction is a process applied to sound level measurements to account for variations in air temperature that can affect sound propagation and the accuracy of readings. As sound travels through air, changes in temperature can influence the speed of sound and consequently alter sound level meter readings, making it essential to apply corrections for accurate assessments.